Iacubovici L, Karol D, Baar Y, Beri A, Herzberg H, Zarour S, Goren O, Cohen B: Assessment of Intra-Abdominal Pressure with a Novel Continuous Bladder Pressure Monitor—A Clinical Validation Study. Life 2023; 13

January 27, 2023

Introduction:
Intra-abdominal hypertension and the resulting abdominal compartment syndrome are serious complications of severely ill patients. Diagnosis requires an intra-abdominal pressure (IAP) measurement, which is currently cumbersome and underused. We aimed to test the accuracy of a novel continuous IAP monitor.
Methods:
Adults having laparoscopic surgery and re- quiring urinary catheter intra-operatively were recruited to this single-arm validation study. IAP measurements using the novel monitor and a gold-standard foley manometer were compared. After anesthesia induction, a pneumoperitoneum was induced through a laparoscopic insufflator, and five randomly pre-defined pressures (between 5 and 25 mmHg) were achieved and simultaneously measured via both methods in each participant. Measurements were compared using Bland–Alt- man analysis.
Results:
In total, 29 participants completed the study and provided 144 distinct pairs of pressure measurements that were analyzed. A positive correlation between the two methods was found (R2 = 0.93). There was good agreement between the methods, with a mean bias (95% CI) of −0.4 (−0.6, −0.1) mmHg and a standard deviation of 1.3 mmHg, which was statistically significant but of no clinical importance. The limits of agreement (where 95% of the differences are expected to fall) were −2.9 and 2.2 mmHg. The proportional error was statistically insignificant (p = 0.85), sug- gesting a constant agreement between the methods across the range of values tested. The percentage error was 10.7%. Conclusions:
Continuous IAP measurements using the novel monitor performed well in the clinical setup of controlled intra-abdominal hypertension across the evaluated range of pressures. Further studies should expand the range to more pathological values.
Keywords:
intra-abdominal hypertension; abdominal compartment syndrome; intra-abdominal pressure monitoring; foley manometer